Front-panel procedures (no software required)
Both procedures are documented in the OEM Operation & Service Manual.
USER PROGRAMS menu - read the PM counters and firmware version
- MODE switch to OFF.
- Press and hold MUTE/CANCEL and MANUAL BREATH/TRIGGER together.
- While holding both, turn MODE to A/C, SIMV, CPAP or CAL.
- The LCD enters USER PROGRAMS. The bottom line of every menu page shows:
#### Hrs- cumulative hours of operation since the last preventive maintenance### Dys- cumulative days since the last preventive maintenanceVersion #.##- firmware version
Write all three down before and after any reset; they are the proof that the reset worked.
Menu items: LCD backlight threshold (SAVE stores it), LCD contrast (SAVE stores it), trigger level sensitivity (temporary), set spontaneous flow (temporary), DEMONSTRATION MODE (training without disconnect alarm / proximal transducer; cancelled at power-off), TEST BACKUP VENTILATOR. Press the button next to EXIT to leave.
MODE = CAL - transducer zero
- Disconnect the patient circuit so the transducer (green) barb is at ambient pressure. Do not occlude anything.
- Turn MODE to CAL. The AMC shows the calibration-in-progress message; wait ~3 s until it
reports complete. Turning the knob away early causes
TRANSDUCER CALIBRATION ABORT. - Turn MODE to the desired mode; the bar graph should sit at 0 cmH2O.
The firmware also re-zeroes automatically during SELF-CHECK and (firmware 1.85X) every 5 minutes (AUTO CAL). After an RTC reset the manual says the unit reboots with a Transducer Calibration alarm and must be put in CAL mode once.
DEMO mode for classroom use
USER PROGRAMS → DEMONSTRATION MODE gives A/C DEMO, SIMV DEMO and CPAP DEMO: settings, alarms and simulated waveforms without gas. It cancels at power-off. It does not clear the PM advisory, which is why the serial reset is needed.
Advisory texts you may see on the message centre
| Text | Meaning | Cleared by |
|---|---|---|
PREVENTATIVE MAINTENANCE DUE - CONTACT CUSTOMER SERVICE (or the short form) |
2000 h or 12 months since the last PM, from the RTC's maintenance clocks | the RTC reset (pm-reset); MUTE only silences the tone |
EXTENDED NON-USE/STORAGE - TEST BEFORE PATIENT USE |
raised at power-up after 6 months of continuous non-use, tracked by the same RTC | running the unit; a stored training unit will show it and it is not a failed reset |
TRANSDUCER CALIBRATION - DO NOT ATTEMPT PATIENT USE / TRANSDUCER CALIBRATION ABORT |
transducer needs zeroing, or CAL was interrupted | MODE → CAL |
VENTILATOR FAILURE DETECTED / FAILURE CODE n |
SELF-CHECK failure; the unit does not cycle and cannot be calibrated until cleared | see the code table below, then power-cycle / MODE → CAL |
Failure codes depend on the firmware version shown in USER PROGRAMS (source: Impact service worksheet Rev G/H):
| Code | Ver 1.83X | Ver 1.85X |
|---|---|---|
| 1 | Transducer Cal Fail | Transducer Cal Fail |
| 2 | No Gas, Compressor | No Gas, Compressor |
| 3 | Total Flow Backup | Excessive Airway Pressure |
| 4 | Excessive Airway Pressure | Memory Check |
| 5 | Memory Check | Exhalation Valve Failure |
| 6 | Exhaust Valve Fail | Excessive Negative Pressure |
| 7 | Excessive Negative Pressure | Run-Time Cal Fail |
What the front panel cannot do
There is no front-panel path to reset the PM clock. The manual's only reset is the PC program over RS-232 (section Computer Calibration, 4.0 "Initializing EEPROM / Resetting the Real Time Clock").